Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Graphtec WR8000: (1 year) CAL VER /5520 DATE: 22-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 19 NUMBER OF LINES: 163 ============================================================================= # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # Sub Graphtec WR8000 Chan. Tests (1 year) CAL VER /5520 # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F V 1.002 ASK+ K X 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Warm-up time: 20 min. - REQUIRED 1.004 DISP Temperature: 18 - 28 degC 1.004 DISP Humidity: <75% 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 OPBR The WR8000 can have either 8 or 16 channels. 1.006 OPBR 1.006 OPBR How many channels does this unit have? 1.007 JMPL VALID_INPUT MEM == 8 1.008 JMPL VALID_INPUT MEM == 16 1.009 JMP 1.006 1.010 LABEL VALID_INPUT 1.011 MATH M[1] = MEM 1.012 MATH MEM = 1 1.013 CALL Sub Graphtec WR8000 Chan. Tests (1 year) CAL VER /5520 1.014 JMP 2.001 1.015 EVAL dummy 2.001 RSLT = 2.002 MATH MEM = 2 2.003 CALL Sub Graphtec WR8000 Chan. Tests (1 year) CAL VER /5520 2.004 JMP 3.001 2.005 EVAL dummy 3.001 RSLT = 3.002 MATH MEM = 3 3.003 CALL Sub Graphtec WR8000 Chan. Tests (1 year) CAL VER /5520 3.004 JMP 4.001 3.005 EVAL dummy 4.001 RSLT = 4.002 MATH MEM = 4 4.003 CALL Sub Graphtec WR8000 Chan. Tests (1 year) CAL VER /5520 4.004 JMP 5.001 4.005 EVAL dummy 5.001 RSLT = 5.002 MATH MEM = 5 5.003 CALL Sub Graphtec WR8000 Chan. Tests (1 year) CAL VER /5520 5.004 JMP 6.001 5.005 EVAL dummy 6.001 RSLT = 6.002 MATH MEM = 6 6.003 CALL Sub Graphtec WR8000 Chan. Tests (1 year) CAL VER /5520 6.004 JMP 7.001 6.005 EVAL dummy 7.001 RSLT = 7.002 MATH MEM = 7 7.003 CALL Sub Graphtec WR8000 Chan. Tests (1 year) CAL VER /5520 7.004 JMP 8.001 7.005 EVAL dummy 8.001 RSLT = 8.002 MATH MEM = 8 8.003 CALL Sub Graphtec WR8000 Chan. Tests (1 year) CAL VER /5520 8.004 JMP 9.001 8.005 EVAL dummy 9.001 JMPL CHART_FEED M[1] < 16 9.002 RSLT = 9.003 MATH MEM = 9 9.004 CALL Sub Graphtec WR8000 Chan. Tests (1 year) CAL VER /5520 9.005 JMP 10.001 9.006 EVAL dummy 10.001 RSLT = 10.002 MATH MEM = 10 10.003 CALL Sub Graphtec WR8000 Chan. Tests (1 year) CAL VER /5520 10.004 JMP 11.001 10.005 EVAL dummy 11.001 RSLT = 11.002 MATH MEM = 11 11.003 CALL Sub Graphtec WR8000 Chan. Tests (1 year) CAL VER /5520 11.004 JMP 12.001 11.005 EVAL dummy 12.001 RSLT = 12.002 MATH MEM = 12 12.003 CALL Sub Graphtec WR8000 Chan. Tests (1 year) CAL VER /5520 12.004 JMP 13.001 12.005 EVAL dummy 13.001 RSLT = 13.002 MATH MEM = 13 13.003 CALL Sub Graphtec WR8000 Chan. Tests (1 year) CAL VER /5520 13.004 JMP 14.001 13.005 EVAL dummy 14.001 RSLT = 14.002 MATH MEM = 14 14.003 CALL Sub Graphtec WR8000 Chan. Tests (1 year) CAL VER /5520 14.004 JMP 15.001 14.005 EVAL dummy 15.001 RSLT = 15.002 MATH MEM = 15 15.003 CALL Sub Graphtec WR8000 Chan. Tests (1 year) CAL VER /5520 15.004 JMP 16.001 15.005 EVAL dummy 16.001 RSLT = 16.002 MATH MEM = 16 16.003 CALL Sub Graphtec WR8000 Chan. Tests (1 year) CAL VER /5520 16.004 JMP 17.001 16.005 EVAL dummy 17.001 LABEL CHART_FEED 17.002 RSLT = 17.003 HEAD {CHART FEED MEASUREMENT} 17.004 DISP Select the DIRECT Y-T mode. 17.004 DISP 17.004 DISP Set the GRID to 10mmfine and CHART SPEED to 10mm/s. 17.004 DISP 17.004 DISP Press START/STOP key to feed the paper. 17.004 DISP 17.004 DISP Stop paper feed when the paper has been fed more than 17.004 DISP 30cm (35 to 40cm). 17.005 MESS Measure the grid lines on the chart paper with a 17.005 MESS transparent scale to measure the paper feed accuracy. 17.005 MESS 17.006 MATH S[1] = "Measurement is 295 to 305." 17.007 EVAL -s S[1] : Does paper measure between 295 and 305? 18.001 MESS 18.002 HEAD {PAPER SKEW} 18.003 DISP Select the DIRECT Y-T mode. 18.003 DISP 18.003 DISP Press the START/STOP key to start paper feed. 18.003 DISP When approximately 30cm of chart has been fed, feed 18.003 DISP another one meter of chart paper. 18.003 DISP Press the START/STOP key to stop the paper feed. 18.004 MESS Measure the distance between the paper edge and 18.004 MESS the grid line at the beginning of the 1m of paper. 18.004 MESS 18.004 MESS Measure the distance between the paper edge and 18.004 MESS the grid line at the end of the 1m of paper. 18.004 MESS 18.004 MESS The difference between these two measurements should be 18.004 MESS less than or equal to 1.5. 18.005 MATH S[1] = "Difference is less than or equal to 1.5." 18.006 EVAL -s S[1] : Is the difference less than or equal to 1.5? 19.001 END